Advanced Instruments Inc. 10 Connections – Optional Intrinsic Safety Barrier: See attached addendum Hazardous Area Operation: When used in conjunc
Advanced Instruments Inc. 11 Span Gas Preparation Caution: Do not contaminate the span gas cylinder when connecting the regulator. Bleed the air
Advanced Instruments Inc. 12 Menu Navigation: The four (4) pushbuttons located on the front of the transmitter operate the micro-processor: ¾ gr
Advanced Instruments Inc. 13 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATE 24.5 C 100 KPA >>> MANUAL RA
Advanced Instruments Inc. 14 Calibration Single Point Calibration: As previously described the galvanic oxygen sensor generates an electrical cu
Advanced Instruments Inc. 15 Recommendation: Zero calibration, see Determining True Zero Offset below, is recommended only for online analyzers pe
Advanced Instruments Inc. 16 taped and securely tightened into the mating male quick disconnect fittings which mate with the female fittings on the
Advanced Instruments Inc. 17 PASSED CALIBRATION OR FAILED CALIBRATION Satisfying users that the zero offset is reasonably ac
Advanced Instruments Inc. 18 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATE 24.5 C 100 KPA >>> CALIBRAT
Advanced Instruments Inc. 19 MAIN MENU AUTO SAMPLE MANUAL SAMPLE CALIBRATE 24.5 C 100 KPA >>> CALIBRATI
Advanced Instruments Inc. 2 Table of Contents 1 Introduction 2 Quality Control Certification 3 Safety 4 Features & Specifications
Advanced Instruments Inc. 20 3.3 PPM AUTO SAMPLING 10 PPM RANGE 24.5 C 100 KPA 25. If the calibration i
Advanced Instruments Inc. 21 6 Maintenance Generally, cleaning the electrical contacts or replacing filter elements is the extent of the maintenan
Advanced Instruments Inc. 22 8 Troubleshooting Symptom Possible Cause Recommended Action Slow recovery At installation, defective s
Advanced Instruments Inc. 23 Symptom Possible Cause Recommended Action Erratic O2 reading or No O2 reading Test sensor indepe
Advanced Instruments Inc. 24 9 Warranty The design and manufacture of the GPR-1535 GB PPM Oxygen Transmitter is performed under a certified Qual
Advanced Instruments Inc. 25 10 MSDS – Material Safety Data Sheet Product Identification Product Name Oxygen Sensor Series - PSR, GPR, AII, XLT
Advanced Instruments Inc. 26 Reactivity Data Stability Stable Conditions Contributing to Instability None Incompatibility KOH = Avoid contact w
Advanced Instruments Inc. 3 2 Quality Control Certification Date: Customer:
Advanced Instruments Inc. 4 3 Safety General This section summarizes the essential precautions applicable to the GPR-1500AIS ppm Oxygen Transmitt
Advanced Instruments Inc. 5 Maintenance Serviceability: Except for replacing the oxygen sensor, there are no parts inside the transmitter for the o
Advanced Instruments Inc. 6 4 Features & Specifications Specifications and pricing are subject to change without notice. See last page for
Advanced Instruments Inc. 7 5 Operation Principle of Operation The GPR-1535 GB incorporates a variety of ppm range advanced galvanic fuel cell ty
Advanced Instruments Inc. 8 Installation Considerations Gas Sample Stream: Ensure the gas stream or composition of the controlled atmosphere of t
Advanced Instruments Inc. 9 Recommendations to avoid erroneous oxygen readings and damaging the sensor: ¾ Do not place your finger over the vent